Below, all the parameter icons with their adjustment options are shown.
Which parameters are available to adjust the programme depends on the appliance for
which a programme is to be created. Only those parameters are available that the appli-
ance is able to implement. For appliances without humidity regulation, for example, no
humidity icon is available. The respective adjustment options (temperature ranges etc.) are
Depiction
in icon bar
Meaning Depiction on
editor thread
Function and
adjustment options
Tolerance band
Maintains a set temperature for a
Name of programme segment
Duration (time or infinite
Temperature to be maintained
Tolerance value above/below
Alarm if limits are exceeded
(For programme example, see page
0,1 K/min
Tolerance band
Increases or decreases temperature
over a specific time to a set value.
Name of programme segment
Target (setpoint) temperature
Tolerance value above/below
(For programme example, see page
When run, this is displayed in the
status bar of the appliance
Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.
Commenti su questo manuale