Depiction
in icon bar
Meaning Depiction on
editor thread
Function and
adjustment options
5.0%rh
Tolerance band
+
-
Alarm
Safe
on off
70.0%rh
12h:00 m
Hold
0.0%rh
Maintains a specific humidity for a
Name of programme segment
Duration (time or infinite
Humidity value to be main-
Tolerance value above/below
Alarm if limits are exceeded
2h:30 m
40.7%rh 70.6%rh
5.0%rh
5.0%rh
+
-
SPWT
on off
Tolerance bandIncrease
Increases or decreases humidity
over a specific time to a specific
Name of programme segment
Target (setpoint) humidity
Tolerance value above/below
When run, this is displayed in the
status bar of the appliance
Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.