Depiction
in icon bar
Meaning Depiction on
editor thread
Function and
adjustment options
+
-
Tolerance band
Alarm
Safe
on off
hold p
20mb
10mb
2h:30 m
500mb
Maintains a specific pressure for a
Name of programme segment
Duration (time or infinite
Pressure to be maintained
Tolerance value above/below
Alarm if limits are exceeded
+
-
Tolerance band
1100mb
decrease
2h:30 m
500mb
10mb
20mb
SPWT
on off
Increases or decreases pressure
over a specific time to a specific
Name of programme segment
Target (setpoint) pressure
Tolerance value above/below
When run, this is displayed in the
status bar of the appliance
Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.